A144

INCOMPETENT PROSTHETIC AORTIC VALVE

Description
This specimen is a Carpentier-Edwards prosthetic aortic valve. There are multiple tears in the valve leaflets immediately adjacent to the free edges. At operation these appeared to be due to impaction of the cusps during systole on the fairly long cut ends of the placement sutures. There was no evidence of an infective process.
History
The patient had undergone aortic and mitral valve replacement three years earlier for aortic and mitral regurgitation secondary to rheumatic fever. Evidence of incompetence of the prosthetic valve was found on routine follow-up examination.